• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNKN TRIGEN INTERTAN INTERTROCH ANTEGR NAIL; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNKN TRIGEN INTERTAN INTERTROCH ANTEGR NAIL; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Catalog Number UNKN02100103
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Failure of Implant (1924)
Event Date 09/01/2020
Event Type  Injury  
Event Description
It was reported that on literature review "effectiveness of proximal femoral nail anti-rotation combined with minimally invasive percutaneous plate osteosynthesis versus intertan intramedullary nail fixation in treatment of intertrochanteric fracture with incomplete lateral wall", one patient suffered from a screw backing out after a primary fixation procedure with the trigen intertan system.It is unknown how the adverse event was treated.No further information is available.
 
Manufacturer Narrative
Internal complaint reference: case (b)(4)."effectiveness of proximal femoral nail anti-rotation combined with minimally invasive percutaneous plate osteosynthesis versus intertan intramedullary nail fixation in treatment of intertrochanteric fracture with incomplete lateral wall".Doi: 10.7507/1002-1892.202001071.
 
Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that without clinically relevant patient-specific supporting documentation, a thorough medical investigation could not be performed.The root cause and/or patient outcome beyond that which was documented in the article could not be confirmed nor concluded.No further medical assessment is warranted at this time.Should additional medical information be provided this complaint will be re-assessed.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to procedural/user error, surgical complication and/ or size of device.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN TRIGEN INTERTAN INTERTROCH ANTEGR NAIL
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14028864
MDR Text Key288778252
Report Number1020279-2022-01586
Device Sequence Number1
Product Code JDG
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Literature,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKN02100103
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-