Manufacturer's investigation conclusion: the centrimag 2nd generation primary console (serial #: (b)(6)) was returned for evaluation with the reported event of the motor speed and flow dropping was confirmed.The 2nd generation primary console (serial #: (b)(6)) was returned for analysis and a log file was downloaded for review.A review of the downloaded log file showed events spanning approximately ((b)(6) 2022 per time stamp).Events occurring on (b)(6) 2022 took place during lab testing at abbott.The console was operating a pump at the speed of 4500 rpm with a flow of 4.9lpm.On (b)(6) 2022 at 15:47:13, sub fault ¿sf_ifd_speed_mismatch¿ activated and the speed dropped to 0 rpm and the flow became a negative value.¿set pump speed not reached: m5¿ and ¿flow below minimum: f3¿ alarms were active.These alarms cleared and the speed and flow were resumed at 15:47:25.This sequence of events occurred again at 15:48:15.The speed and flow was maintained until the console was powered down on (b)(6) 2022 at 17:26:48.There were no other notable alarms active in the log file.The centrimag 2nd generation primary console was returned to the service depot.The console was connected to a test loop and was run for several days.No alarms were observed.A full functional test was performed, and all tests were passed.The reported event was unable to be correlated to an issue with the returned console.The device history records were reviewed for the centrimag 2nd generation primary console (serial #: (b)(6)) and the console was found to pass all manufacturing and qa specifications.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." the 2nd generation centrimag system operating manual table 16 entitled ¿console alarms & alerts¿ addresses how to properly interpret and troubleshoot all system alarms including m5 and f3 alarms.No further information was provided.The manufacturer is closing the file on this event.
|