Model Number A2114
Device Problem Mechanical Problem (1384)
Patient Problem Laceration(s) (1946)
Event Date 03/18/2022
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that the rocker arm on the mayfield modified skull clamp (a1059) came unlocked prior to posterior cervical fusion procedure during patient positioning.As a result, the patient sustained a 9cm laceration.There was a 20 minute delay in surgery as a new mayfield had to be applied and the surgery proceeded as scheduled.No additional information was provided.
 
Manufacturer Narrative
The mayfield infinity xr2 skull clamp (a2114) was returned for evaluation.Device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the investigation of the returned device showed that the reported complaint was confirmed from the evaluation.Unit received had the set screw replaced due to a loose index knob.Both the right and left pawls were replaced due to wear.General maintenance performed.Root cause - complaint confirmed via inspection of the unit.The index knob was loose and had the set screw replaced.Unit required general maintenance and replacement of worn components; probable root cause is routine use and wear.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key14048018
MDR Text Key291208961
Report Number3004608878-2022-00061
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253761
UDI-Public10381780253761
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberA2114
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-