Model Number A3059
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Mayfield composite series skull clamp (a3059) was returned for evaluation: device history record (dhr): the dhr shows no abnormalities related to the reported failure.Failure analysis: the investigation of the returned device found no device deficiencies that would have contributed to the reported complaint.The unit received passed all specific functional testing.Repairs could not duplicate the complaint.Unit will be returned to the customer without requiring any repairs.Root cause analysis: evaluation found no device deficiencies that would have contributed to the reported complaint.Repairs could not duplicate complaint, unit passes all specific functional testing.
 
Event Description
A facility reported that when the swivel lock on the mayfield composite series skull clamp (a3059) was engaged and the swivel arm experiences sideways pressure, they were unable to unlock the swivel arm in order to remove the skull clamp from the patient.No patient injury or delay in surgery was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD COMPOSITE SERIES SKULL CLAMP
Type of Device
¿COMPOSITE SERIES¿
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key14048022
MDR Text Key298465009
Report Number3004608878-2022-00060
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253792
UDI-Public10381780253792
Combination Product (y/n)N
PMA/PMN Number
K142238
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA3059
Device Catalogue NumberA3059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2022
Date Manufacturer Received03/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-