C.R. BARD, INC. (COVINGTON) -1018233 LUBRI-SIL FOLEY CATHETERS; UNKNOWN FOLEY CATHETERS
|
Back to Search Results |
|
Catalog Number UNKNOWN |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Pain (1994); Urinary Tract Infection (2120); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 03/16/2022 |
Event Type
Injury
|
Event Description
|
It was reported that the customer mentioned about bacteriuria, symptomatic urinary tract infection, pain, urethral or bladder trauma, catheter blockage, leakage and bladder stone(s) when they were asked about the complications while using the lubrisil hydrogel coated silicone foley catheter.It was unknown what medical intervention was provided.
|
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
|
|
Manufacturer Narrative
|
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be due to "tortuous anatomy".It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage silicone and may cause the balloon to burst.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Unless package is opened or damaged.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Single use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip tip syringe.Do not use needle.Recommended inflation capacities 3cc balloon: use 5cc sterile water 5cc balloon: use 10cc sterile water 30cc balloon: use 35cc sterile water do not exceed recommended capacities.Note: aggressive traction, particularly in the presence of suturing, is not recommended for silicone foley catheters.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.H11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
Event Description
|
It was reported that the customer mentioned about bacteriuria, symptomatic urinary tract infection, pain, urethral or bladder trauma, catheter blockage, leakage and bladder stone(s) when they were asked about the complications while using the lubrisil hydrogel coated silicone foley catheter.It was unknown what medical intervention was provided.
|
|