• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC VENTIPAC; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC VENTIPAC; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Catalog Number 130003-00
Device Problem False Alarm (1013)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2022
Event Type  malfunction  
Event Description
It was reported that there were high pressure alarms while the gauge read zero.No patient involvement was reported.
 
Event Description
Additional information was received indicating that the exact date of the event is unknown, but believed to have occurred somewhere around 22-february-2022.Confirmation that no patient was involved.
 
Manufacturer Narrative
Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found minor wear and tear on the housing, no other apparent physical damages found.Functional testing found the high pressure led illuminates at all times due to faulty led printed circuit board (pcb).The root cause of the reported issue was found to be component failure.Replaced mri battery, o-rings and input filter for preventive maintenance.Replaced faulty led pcb to resolve the customer reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC VENTIPAC
Type of Device
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
bramingham business park, enterprise way
luton, bedforshire LU3 4 BU
UK  LU3 4BU
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
bramingham business park, enterprise way
luton, bedforshire LU3 4 BU
UK   LU3 4BU
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14090514
MDR Text Key289187929
Report Number3012307300-2022-06257
Device Sequence Number1
Product Code BTL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number130003-00
Was Device Available for Evaluation? No
Date Returned to Manufacturer03/18/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/26/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-