• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DENALI JUGULAR FILTER; VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DENALI JUGULAR FILTER; VENA CAVA FILTER Back to Search Results
Model Number DL900J
Device Problems Failure to Advance (2524); Device Dislodged or Dislocated (2923); Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2022
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 02/2025).
 
Event Description
It was reported that during a jugular filter placement procedure, the device allegedly deployed in sheath.The procedure was completed using another device.There was no reported patient injury.
 
Event Description
It was reported that during a filter placement procedure, the device allegedly detached from the pusher rod and failed to advance.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one denali jugular filter kit was retuned for evaluation.The introducer sheath, filter storage tube and touhy-borst adapter were received in a connected condition.The pusher catheter was loaded within the touhy-borst adapter, filter storage tube and introducer sheath.The introducer sheath was x-rayed and it was noted that the filter was within the sheath and filter hook was disengaged from the pusher gripper.The pusher catheter was used to deploy the filter from the sheath; all limbs are present and uncrossed.Therefore, the investigation is confirmed for the device dislodged or dislocated and filter advancement issue as the filter was within the introducer sheath and filter hook was disengaged from the pusher gripper.A definitive root cause for the reported device dislodged and failure to advance issues are could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 02/2025), g3.H11: b5, h6 (method, result).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failures as no objective evidence was provided for review.A definitive root cause for the reported device dislodged and failure to advance issues are could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 02/2025), g3, h6 (method).H11: g1, h6 (device, result, conclusion).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event Description
It was reported that during a jugular filter placement procedure, the device allegedly deployed in sheath.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DENALI JUGULAR FILTER
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14100375
MDR Text Key289176469
Report Number2020394-2022-00289
Device Sequence Number1
Product Code DTK
UDI-Device Identifier00801741040795
UDI-Public(01)00801741040795
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDL900J
Device Catalogue NumberDL900J
Device Lot NumberGFGN1019
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/26/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
Patient Weight80 KG
-
-