Model Number OER-PRO |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/17/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The biomed later called to inform olympus he was able to resolve the issue himself and to cancel the service request for a field service engineer (fse).The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
|
|
Event Description
|
The customer reported to olympus that since the field service engineer was last onsite their usage display category does not show any counters on their endoscope reprocessor.There was no patient involvement.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the event was caused by the memory of circuit board malfunctioned or the user set the counter to ¿-¿ but misunderstood it as an error.The event can be detected by following the instructions for use which state: "operation manual setting the disinfection cycle and day counter set [- -] if you do not want to set the disinfection cycle count.Set [- -] if you do not want to set the day counter.¿ olympus will continue to monitor field performance for this device.
|
|