• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG METAL METAL BALL HEAD 28S CONE 12/14; PROSTHESIS, HIP, SEMI-CONSTRAINED (METAL CEMENTED ACETABULAR COMPONENT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG METAL METAL BALL HEAD 28S CONE 12/14; PROSTHESIS, HIP, SEMI-CONSTRAINED (METAL CEMENTED ACETABULAR COMPONENT) Back to Search Results
Catalog Number 75004135
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Pain (1994)
Event Date 09/22/2006
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Event Description
It was reported that, after a left tha had been performed on (b)(6) 2005, the patient experienced unexpected pain.This event was addressed via revision surgery on (b)(6) 2006 to explant cup, liner, head and stem.Patient health status is not known.This information was provided by the national joint registry of (b)(6), following a review of the data conducted in september 2021.
 
Manufacturer Narrative
H3, d4, h6: a total of 84 revision surgeries involving bicon-plus devices was reported by the uk national joint registry.140 complaints were opened accordingly.A review of the device history record (dhr), of the complaint history (chr), of the instruction for use (ifu), of the risk management file (rmr) and of past escalation (per) was performed.No relevant supporting clinical information was provided.Based on insufficient information, a thorough clinical assessment cannot be performed at this time.The patient's current condition is unknown and the patient impact beyond the reported events could not be determined.Based on the chr and dhr evaluation performed there is no indication that any failure to meet specification contributed to the reported events.The reported failure modes are outlined in the device instruction for use lit no 12.23 03/21.The occurrence and severity of the reported issues are in line with the risk plans.The root cause of the reported events was attributed to known inherent risks of the devices.No corrective actions have been initiated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METAL METAL BALL HEAD 28S CONE 12/14
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED (METAL CEMENTED ACETABULAR COMPONENT)
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14132858
MDR Text Key289421281
Report Number9613369-2022-00179
Device Sequence Number1
Product Code JDL
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/29/2011
Device Catalogue Number75004135
Device Lot Number0412.10.1631
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/30/2004
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention; Other;
Patient Age70 YR
Patient SexFemale
-
-