Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient with lumbar disc herni ation for lumbar intervertebral decompression, fusion and fixation spinal therapy.It was reported that the device has slipped.There was no patient symptoms reported.Screw plug thread was worn, deformed and bent.
|
H3: product analysis #281561158:part # 5540030 ; lot # h5673395 visual and optical inspection did not reveal any damage to the female torx head, but the threads of the set screw appear to be worn/used.Functional check with a sample bone screw confirmed the set screw was able to thread into the head of the screw without any issue.The bottom node of the set screw does not show signs of off axis/ uneven wear from the seating of the rod.The set screw appears to function as intended.No fault found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|