Model Number 4470 |
Device Problems
High impedance (1291); Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/28/2022 |
Event Type
Injury
|
Event Description
|
It was reported that this right atrial (ra) lead was explanted and replaced due to high out-of-range pace impedance measurements greater than 3,000 ohms.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
The local area field representative was contacted for additional information.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
The local area field representative was contacted for additional information.If information is provided in the future, a supplemental report will be issued.-- if information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that this right atrial (ra) lead was explanted and replaced due to high out-of-range pace impedance measurements greater than 3,000 ohms.No additional adverse patient effects were reported.Additional information received reported that during the lead revision, the leader/header connections were checked.When the lead was tested with the pacing system analyzer (psa), the pace impedance measurement was 500 ohms and within normal limits.It was unclear whether the connection had contributed to the high out-of-range pace impedance measurements, however the decision was made to extract the lead.The pacemaker remains in service.It was noted that the patient was doing fine.No additional adverse patient effects were reported.
|
|
Search Alerts/Recalls
|