• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problems No Display/Image (1183); Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2022
Event Type  malfunction  
Event Description
It was reported that the external pulse generator (epg) was unable to sense and was displaying inaccurate readings.The epg has been returned to service.There was no patient involvement.
 
Manufacturer Narrative
Product analysis: the analysis could not confirm the customer comment of the external pulse generator (epg) was unable to sense and was displaying inaccurate readings.It was determined at analysis that the hanger assembly and upper case were broken.The atrial encoder threads were stripped.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key14165947
MDR Text Key289738214
Report Number2182208-2022-01261
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00763000162443
UDI-Public00763000162443
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/13/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-