Model Number 93330
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Impaired Healing (2378); Fluid Discharge (2686); Drug Resistant Bacterial Infection (4553)
Event Type  Injury  
Manufacturer Narrative
This report is submitted on april 22, 2022.
 
Event Description
Per the clinic, the patient experienced an infection (b)(6) (treatment unknown) and drainage at the implant site.Initially the abutment was removed but the infection did not resolve.The device was explanted on (b)(6) 2022.It is unknown if there are plans to re-implant the patient with another device.
 
Manufacturer Narrative
Per the clinic, the patient was placed under general anaesthesia in order to have the implant explanted.The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Adverse skin reactions around the baha abutment, ranging from slight redness to infected tissue, are common complications with baha implants, and can occur at any time following implantation.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.This report is submitted on june 13, 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIA400 IMPLANT 4MM W ABUTMENT 8MM
Type of Device
COCHLEAR BAHA CONNECT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
po box 82
mölnlycke, 435 3 3
SW  435 33
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW   435 33
Manufacturer Contact
debbie ang
unit ug-1, vertical podium
no.8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key14179074
MDR Text Key289865661
Report Number6000034-2022-01118
Device Sequence Number1
Product Code MAH
UDI-Device Identifier09321502022705
UDI-Public(01)09321502022705(10)134810(17)180430
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number93330
Device Catalogue Number93330
Device Lot Number134810
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/18/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-