Catalog Number 07027559190 |
Device Problems
Incorrect Measurement (1383); High Test Results (2457)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/28/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Qc was within range on the day of the event.On (b)(6) 2022 the customer performed dilution testing on the sample and the results from the e801 unit were: 1:2 dilution result >1000 uu/ml; 1:4 dilution result >1000 uu/ml; 1:8 dilution result >1000 uu/ml; 1:16 dilution result 875 uu/ml with a final result of 14000 uu/ml; 1:32 dilution result 465 uu/ml with a final result of 14880 uu/ml; 1:64 dilution result 237 uu/ml with a final result of 15168 uu/ml; 1:128 dilution result 113 uu/ml with a final result of 14464 uu/ml.On (b)(6) 2022 the customer treated the sample by polyethylene glycol precipitation (peg) to test for an interference and the insulin result from the e801 unit was >1000 uu/ml.The customer took a different patient sample to see if peg treatment worked.The initial insulin result was 39.8 uu/ml; the result after peg was 19.0 uu/ml.A new sample was obtained from the patient on (b)(6) 2022 and the anti-insulin, anti-islet cell and anti-glutamate decarboxylase antibodies results were all negative.The patient was retested multiple times both in a fasting state and postprandial for insulin, glu and cp with the following results: the insulin results from the e801 unit were: fasting:>1000 uu/ml normal reference range (2.6-24.9 uu/ml); after meal 0.5 h: >1000 uu/ml; after meal 1.0 h: >1000 uu/ml; after meal 2.0 h: >1000 uu/ml; after meal 3.0 h: >1000 uu/ml.The cp results from the e801 unit were: fasting: 10.87 nmol/l normal reference range (0.37-1.47 nmol/l); after meal 0.5 h: 11.33 nmol/l; after meal 1.0 h: 11.17 nmol/l; after meal 2.0 h: 10.7 nmol/l; after meal 3.0 h: 11.73 nmol/l.The glu results from an unspecified, non-roche method were: fasting: 3.8 nmol/l normal reference range (3.9-6.1nmol/l); after meal 0.5 h: 10.7 nmol/l; after meal 1.0 h: 15.3 nmol/l normal reference range (6.7-11.1 nmol/l); after meal 2.0 h: 20.5 nmol/l normal reference range (3.9-7.8 nmol/l); after meal 3.0 h: 16.3 nmol/l normal reference range (3.9-7.8 nmol/l).The patient sample was requested for investigation.
|
|
Event Description
|
The initial reporter complained of a discrepant high result for 1 patient sample tested for elecsys insulin (insulin) on a cobas e 801 analytical unit compared to the siemens method.The 2 hour post-prandial result from the e801 unit was >1000 uu/ml.The sample was tested by the siemens method with a result of 101 uiu/ml.The questionable result was reported outside of the laboratory.The e801 unit serial number was (b)(4).
|
|
Manufacturer Narrative
|
Calibration signals from (b)(6) 2022 were within specification.The sample was not available for further investigation.A general reagent issue can be excluded as the quality validation data provided was within the expected range.The investigation did not identify a product problem.The cause of the event could not be determined.
|
|
Manufacturer Narrative
|
Section a2, age was updated.On (b)(6)-2022 a new sample from the patient was tested.The fasting insulin result from the e801 module was >1000 uu/ml.The repeat test was >1000 uu/ml.The patient's fasting c-peptide result was 6.77 nmol/l.These results were reported outside of the laboratory to the same doctor from the initial event that was familiar with the patient.Calibration and qc were acceptable.No instrument issues were identified.The sample was requested for investigation, however, the patient sample is no longer available for further investigation.A general reagent problem can be excluded as the quality validation data provided was within expectations.The high insulin results are consistent with higher insulin production; this is supported by the elevated c-peptide result and the presence of anti-insulin antibodies recognized by the elecsys insulin assay.The investigation did not identify a product problem.The cause of the event could not be determined.
|
|
Search Alerts/Recalls
|