The customer reported that her olympus hf monopolar cable began smoking during procedure preparation.According to the initial reporter, the intended procedure type is unknown.However, she did confirm that the staff completed the case without any patient harm by using another cable.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the lot number was not provided.However, the manufacturing and quality control review was performed for the last 24 months of production and there are no non-conformities or deviations regarding the described issue.Based on the results of the investigation, it is very likely that the age-related wear in connection with repeated extreme bending or tensile loads most likely led to the breakage of single or all the wires in the cable.This can cause a voltage spike at the damaged area when the generator is activated which may result in a spark and complete disconnection of the plug.The specific cause of the event was likely due to age-related wear in connection with improper handling.The service life of the cable is limited to 12 months.After this time, the cable should no longer be used.The following information is stated in the instructions for use regarding proper handling of the device: "additionally, the cable must be checked for damage before each use and after reprocessing.By gently pulling on the plug (max.20nm), it can be determined whether the copper strand of the cable is already damaged.If the cable does not give way but remains rigid, the cable is most likely intact.In order not to shorten the service life of the cable any further, the cable should not be wound up with a loop diameter of less than 10cm and when pulling out the cable, the plug should be pulled and not the cable.".Olympus will continue to monitor field performance for this device.
|