Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having screw replacement and extension.It was reported that the tip was broken when the screw for replacement was being removed.The debris was collected with a tweezer.No residue in the body.A lot of cleaning was performed just in case.The screw was removed successfully with another driver, and the screw was replaced with a new one.There was procedural delay for less than 60 mins.The screw was replaced because the l5 screw was loose.Also, a fixed extension was performed due to a th11 compression fracture.Even though the looseness of the screw in the l5 was observed in the surgery, it was not the cause of the fracture.Therewere no further complications reported regarding the event.
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|