The initial reporter stated they received questionable results for one patient sample tested with gluc3 glucose hk gen.3 on two cobas pro c503 modules.The sample initially resulted in a glucose value of 3.68 mmol/l on the first c503 system (serial number (b)(4)) and this vale was reported outside of the laboratory to the clinician.The clinician asked for a new sample to be collected from the patient and tested as the result was not consistent with the patient's clinical status and the patient had a capillary glucose result of 1.2 mmol/l.The sample was repeated on a second c503 analyzer (serial number unknown) and the result was identical to the first value.A second sample collected from the patient resulted in a glucose value of 1.0 mmol/l when tested on the first c503 system.A new capillary sample collected from the patient resulted in a glucose value of 1.1 mmol/l.The complained sample was later repeated on the first c503 system on (b)(6) 2022, resulting in a glucose value of 3.99 mmol/l.The glucose reagent lot number and expiration date used for the first measurement were requested, but not provided.The second measurement of the sample performed on (b)(6) 2022 was tested using glucose reagent lot 614042.The reagent expiration date was requested, but not provided.
|