Manufacturing device history record review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found a cracked tank cover.Wear and tear damaged enclosure, front cover, and pole clamp.A faded line cord, outdated printed circuit board (pcb) and power switch.Functional testing found the f1 fuse on the pcb shorted out confirming the customer complaint.The root cause of the reported issue was found to be a blown fuse.No action taken due to the age and condition of the device.It was deemed beyond economical repair and will be scrapped.
|