• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VIGILANT EL ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VIGILANT EL ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number D233
Device Problems Pocket Stimulation (1463); Migration (4003)
Patient Problems Device Overstimulation of Tissue (1991); Twiddlers Syndrome (4563)
Event Date 03/03/2022
Event Type  Injury  
Event Description
It was reported that the patient with this implantable cardioverter defibrillator (icd) felt that the device was kicking and needs replacement or at least fix the lead.Additional information from the field representative was obtained which indicated that the patient was experiencing cardiac stimulation and the chest-xray showed that the device was displaced with multiple rotations of leads indicating twiddler's syndrome.A system revision was performed.The device remains in service.No additional adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIGILANT EL ICD DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14230335
MDR Text Key290257727
Report Number2124215-2022-13724
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526587801
UDI-Public00802526587801
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960040/S385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/19/2023
Device Model NumberD233
Device Catalogue NumberD233
Device Lot Number609857
Was Device Available for Evaluation? No
Date Manufacturer Received03/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/27/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age67 YR
Patient SexFemale
-
-