• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD RESPIRATORY HUMIDIFIER; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD RESPIRATORY HUMIDIFIER; BTT Back to Search Results
Model Number MR850
Device Problems No Audible Alarm (1019); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint mr850 respiratory humidifier is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow-up report upon completion of our investigation.
 
Event Description
A healthcare facility in (b)(6) reported that the audible alarm of a mr850 respiratory humidifier was not functioning.There was no reported patient involvement.
 
Manufacturer Narrative
(b)(4).Corrected data: section d9 device available for evaluation? no.Section h3 device evaluated by manufacturer? other.Method: the complaint mr850 respiratory humidifier was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information provided by the customer and our knowledge of the product.Results: the customer reported that the audible alarm of a mr850 respiratory humidifier was not functioning.Conclusion: without the complaint device, we are unable to determine what may have caused the reported failure.Our mr850 product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking and performance testing of the mr850 heater base.In addition, the product technical manual states that "all servicing procedures shall be followed by a humidifier functional test and an electrical safety test, to ensure proper operation".The mr850 is equipped with visual alarm indicators in addition to the audible alarm.
 
Event Description
A healthcare facility in new jersey reported that the audible alarm of a mr850 respiratory humidifier was not functioning.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRATORY HUMIDIFIER
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine 92618
9494534000
MDR Report Key14231873
MDR Text Key294658826
Report Number9611451-2022-00406
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR850
Device Catalogue NumberMR850
Device Lot Number080513
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/13/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-