MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 ENDO-FLO IRRIGATOR; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 ENDO-FLO IRRIGATOR; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 0037970
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2022
Event Type  malfunction  
Manufacturer Narrative
As reported, the bard/davol endo-flo irrigator leaks fluid when used with a reusable metal non-bard/davol trumpet valve instrument, and there is no leaking when the endo-flo irrigator is being used with disposable plastic instruments.A photo was provided that confirms a fluid leak at the luer lock location where the reusable metal non-bard/davol trumpet valve instrument connects to the irrigation tubing of the bard/davol endo-flo.Neither the bard/davol endo-flo, or the reusable metal non-bard/davol trumpet valve instrument has been provided for evaluation; therefore the potential cause of the incompatibility/leaking between the two devices could not be determined.No lot number has been provided; therefore, a review of the manufacturing records is not possible.
 
Event Description
As reported, during an unknown procedure, the bard/davol endo-flo irrigator leaked irrigation fluid when connected to the luer-lock of the non-bard/davol metal handpiece.As reported, connector did not seal and it appeared "like the threads are of the wrong pattern".There was no reported patient injury.Reporter stated "the tubing does not leak with disposable instruments.It only leaks with the metal instrument.It does not leak when connecting to a plastic connector.When attaching the tubing to the metal irrigation instrument, there is significant leakage at the connector.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO-FLO IRRIGATOR
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
DAVOL SURGICAL INNOVATIONS -9616067
ave. roberto fierro #6408
parque industrial aeropuerto
cd. juarez, chih s.a. de c.v. 32690
MX   32690
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key14238065
MDR Text Key290294559
Report Number1213643-2022-00199
Device Sequence Number1
Product Code HET
UDI-Device Identifier00801741065217
UDI-Public(01)00801741065217
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K902722
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0037970
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-