Additional information was provided in sections d.9, h.3, h.6 and h.10.The company service representative examined the system and was unable to confirm or replicate the reported event.As a preventative measure, the host, and the multifunction input output (mfio) printed circuit board (pcb) were replaced.The system was then tested and met all product specifications.The host was received, and a visual assessment of the returned sample revealed no obvious nonconformities.The host sample was installed into a calibrated console and there were no advisory messages during startup.The sample passed all functional tests, and the reported event could not be replicated, and the host was working as intended.A system manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.The system was found to meet specifications; therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|