• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA PLUS VASCULAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA PLUS VASCULAR COVERED STENT Back to Search Results
Catalog Number AASME10100
Device Problems Loss of or Failure to Bond (1068); Break (1069); Positioning Failure (1158)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/18/2022
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The catalog number identified has not been cleared in the us but is similar to the covera plus vascular covered stent products that are cleared in the us.The pro code and 510 k number for the covera plus vascular covered stent products are identified in this report.(expiry date: 07/2023).
 
Event Description
It was reported that during a stent placement procedure, the stent was allegedly broken.It was further reported that stent failed to deploy.The procedure was completed using an another device.There was no reported patient injury.
 
Event Description
It was reported that during a stent placement procedure, the stent was allegedly broken.It was further reported that stent failed to deploy.The procedure was completed using an another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the delivery system was returned for evaluation and the slide block, a force transmitting component was no longer connected to the proximal sheath.The disconnection of the slide block from the proximal sheath, can also be matched with the reported cracking sound.It is considered that the disconnection of the slide block led to the reported impossibility to deploy the stent and the freewheeling effect reported by the customer.Based on evaluation of the sample the inability of an adhesive joint (slide block/tether/diving sheath) to withstand tension force during deployment is confirmed.A definitive root cause could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.Based on the instructions for use supplied with this product delivery system specific events that could be associated with clinical complications include but are not limited to failure to deploy and high deployment forces.Regarding correct deployment the instructions for use states: "(.) maintain a stationary hold on the white stability sheath during covered stent deployment (¿).Hold the white stability sheath as close as possible to the introducer without touching the dark brown moving catheter of the distal catheter assembly.Maintain the remainder of the white stability sheath (.) relaxed and avoid tension." regarding preparation the instructions for use states: "pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated." h10: the catalog number identified in section d4 has not been cleared in the us but is similar to the covera plus vascular covered stent products that are cleared in the us.The pro code and 510 k number for the covera plus vascular covered stent products are identified in d2 and g4.H10: d4 (expiry date: 07/2023), g3, h6 (device).H11: g1, h6 (result).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVERA PLUS VASCULAR COVERED STENT
Type of Device
VASCULAR COVERED STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14272641
MDR Text Key290690852
Report Number9681442-2022-00138
Device Sequence Number1
Product Code PFV
UDI-Device Identifier00801741136047
UDI-Public(01)00801741136047
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
P170042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAASME10100
Device Lot NumberANFU4110
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/26/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-