• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RELIANCE 4-FRONT; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RELIANCE 4-FRONT; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 0676
Device Problems Pocket Stimulation (1463); High Capture Threshold (3266)
Patient Problem Twiddlers Syndrome (4563)
Event Date 03/07/2022
Event Type  Injury  
Manufacturer Narrative
This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.
 
Event Description
It was reported that the patient with this right ventricular (rv) lead experienced pocket stimulation.This lead was also exhibited high pacing threshold measurements.A chest x-ray showed that device and leads was displaced with multiple rotations of leads indicated twiddler syndrome.This lead was explanted and replaced.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIANCE 4-FRONT
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14291118
MDR Text Key290965875
Report Number2124215-2022-14165
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526497001
UDI-Public00802526497001
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P910073/S145
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/18/2022
Device Model Number0676
Device Catalogue Number0676
Device Lot Number504018
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/18/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age67 YR
Patient SexFemale
-
-