Information was received from a healthcare provider via a manufacturer representative regarding a patient having preoperative diagnosis for reinforcement surgery due to bilateral rod breakage.Initial surgery or the procedure taken was t9/s2 fixation with the degenerative kyphoscoliosis it was reported that due to broken rods on both sides, reinforcement was performed using a connector from another company.Although it is not a broken piece, the broken rods on both sides are left as is and reinforcement is performed with connectors.Initial surgery was performed on (b)(6)2020.There were no symptoms reported.There were no further complications reported regarding the event.The broken rods are still in use as they are.They were broken but were not separated into plural pieces/fragments.So, no fragments are remaining inside the patient body.The report indicates 'broken' but the actual condition of the rods seems to be closer to 'having cracks'.There is no plan to remove them.
|
510k number: the device p/n 1556300500 is similar to the device manufactured in us p/n 1556000500.Hence, mentioned 510k license number k111942 is of p/n 1556000500.Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|