• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SCORPION-MULTIFIRE NEEDLE; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SCORPION-MULTIFIRE NEEDLE; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE Back to Search Results
Model Number SCORPION-MULTIFIRE NEEDLE
Device Problems Break (1069); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2021
Event Type  malfunction  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event Description
On (b)(6) 2021 it was reported by a sales representative via sems that during a labrum repair, the surgeon was attempting to pass a ar-4068-25trf suture lasso through capsule/anterior labrum it bent and the tip of two ar-6068c suture lassos broke along with a ar-13995n scorpion needle.The surgeon completed the case with a new scorpion needle to pass sutures, and the broken fragments were removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCORPION-MULTIFIRE NEEDLE
Type of Device
INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key14310175
MDR Text Key295048867
Report Number1220246-2022-04849
Device Sequence Number1
Product Code MDM
UDI-Device Identifier00888867015951
UDI-Public00888867015951
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSCORPION-MULTIFIRE NEEDLE
Device Catalogue NumberAR-13995N
Device Lot Number13520468
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/29/2021
Date Device Manufactured06/14/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-