• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION; VITAL SIGNS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION; VITAL SIGNS MONITOR Back to Search Results
Device Problems Overheating of Device (1437); Temperature Problem (3022)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/15/2022
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer (bme) reported that the nurse noticed that the ac cradle was hot to the touch and took the unit out of service.It was not charging the bsm-1700s being placed in it, and the internal pcb board looked burned.The bedside monitor itself was not returned, but the sc-170r ac cradle used with the bedside monitor was returned.Not in patient use.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.Additional device information: concomitant medical device: the following devices were being used in conjunction with bsm unit and was the unit that failed: ac cradle: model: sc-170r, sn: (b)(4).Device manufacturer date: 02/05/2016.Unique identifier (udi) #: (b)(4).Returned to nihon kohden: 05/03/2022.
 
Event Description
The biomedical engineer (bme) reported that the nurse noticed that the ac cradle was hot to the touch and took the unit out of service.It was not charging the bsm-1700s being placed in it, and the internal pcb board looked burned.The bedside monitor itself was not returned, but the sc-170r ac cradle used with the bedside monitor was returned.Not in patient use.
 
Manufacturer Narrative
Details of complaint: the biomedical engineer (bme) reported that the ac cradle was hot to the touch, and it was no longer charging the bsm-1700s.The internal pcb board looked burned.The sc-170r ac cradle was returned for an exchange, but no bsms were returned.The device was not in patient use.Investigation summary: during the evaluation of the reported device, nihon kohden repair center (nk rc) confirmed the reported complaint and found that the unit was unrepairable.The customer reported that the pcb board for the device was found to be burned.Some common reasons for a pcb board burnout include: overheating: excessive heat can cause the pcb to burn.This can happen if the electronic components on the board generate more heat than the board or its cooling system can dissipate.Overheating can lead to component failure and damage to the pcb.Overcurrent or short circuits: high current levels beyond the rated capacity of the pcb or unexpected short circuits can cause excessive current flow, leading to overheating and burning.This can be caused by faulty components, damaged traces, or incorrect wiring.Voltage spikes: sudden spikes in voltage, often caused by power surges or electrical faults, can overwhelm the protective features of the pcb and cause components to burn.Component failure: individual electronic components on the pcb, such as resistors, capacitors, or integrated circuits, can fail due to manufacturing defects, wear and tear, or exposure to conditions beyond their specifications.A failed component may generate excessive heat and cause the pcb to burn.Environmental factors: exposure to extreme environmental conditions, such as high humidity, corrosive substances, or contaminants, can degrade the pcb's materials over time, potentially leading to burnout.Improper handling: rough handling during assembly, maintenance, or transportation can cause physical damage to the pcb, including broken traces or damaged components, which can contribute to burnout.Moisture and water damage: exposure to moisture or water can cause short circuits and corrosion on the pcb, leading to burnout.Water damage can be especially problematic because it can create conductive paths where they shouldn't exist.Aging: over time, the materials used in the pcb can degrade, and solder joints may develop hairline cracks.This aging process can increase the risk of burnout.Attempt #1 04/18/2022 emailed customer via microsoft outlook for all items under the no information section.The customer responded with complaint details and only the model series number (bsm-1700 series), but did not provide all the requested information.Additional device information: d10: concomitant medical device: the following device, which is the device that experienced the falure and was returned to nk, was used in conjunction with the bsm unit: ac cradle model: sc-170r.Sn: (b)(6).Device manufacturer date: 02/05/2016.Unique identifier (udi) #: (b)(4).Returned to nihon kohden: 05/03/2022.
 
Event Description
The biomedical engineer (bme) reported that the ac cradle was hot to the touch, and it was no longer charging the bsm-1700s.The internal pcb board looked burned.The sc-170r ac cradle was returned for an exchange.The device was not in patient use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NI
Type of Device
VITAL SIGNS MONITOR
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key14312569
MDR Text Key299757603
Report Number8030229-2022-02833
Device Sequence Number1
Product Code MHX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080342
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 11/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
SC-170R SN (B)(6); SC-170R SN (B)(6)
-
-