Model Number 866199 |
Device Problems
Image Display Error/Artifact (1304); Failure to Power Up (1476)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Reporting address state: (b)(6).Reporting address postal: (b)(6).Reporting institution phone #: (b)(6).Reporter phone: (b)(6).A follow up report will be submitted upon completion of the investigation.
|
|
Event Description
|
It was reported to philips that the display is not working.There was no patient involvement.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model # 861290) and will be reported in the united states under device model # 861290.
|
|
Manufacturer Narrative
|
Fse evaluated the device during their initial visit and confirmed the customer complaint that the device had malfunctioned.Fse diagnosed the device and determined processor and power printed circuit board (pcb) were defective.Fse initially replaced defective power pcb and processor however the device malfunction was not resolved.Fse reinstalled original power pcb and processor.Fse secondary diagnosis was input/output (i/o) pcb was the source of the failure.Next the fse replaced the i/o pcb, however the unit was still malfunctioning.Final diagnosis of the fse was that the system on module (som) pcb was defective.Finally replacing the som pcb allowed the unit passed testing and the device was returned to use.Parts were not returned for failure investigation (fi).Trends for this type of issue will continue to be monitored to determine if additional actions are required.The complaint will be processed for closure.If additional information becomes available at a later date, the complaint will be reopened, and a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|