• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE); PROSTHESIS,HIP,HEMI-,TRUNNIONBEARING,FEMORAL,METALPOLYACETAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE); PROSTHESIS,HIP,HEMI-,TRUNNIONBEARING,FEMORAL,METALPOLYACETAL Back to Search Results
Device Problem Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Loss of Range of Motion (2032); Inadequate Osseointegration (2646); Implant Pain (4561)
Event Date 03/26/2021
Event Type  Injury  
Event Description
It was reported that on literature review "can hip resurfacing be safely revised with short-stem total hip arthroplasty? a case series of six patients", 1 patient operated in 2004 with bhr presented limited hip mobility and pain at hip rotation since 2016.After diagnosis with acetabular loosening and acetabular bone defect, a revision using competitor's device was performed.At last study follow up, the patient's outcome was excellent.A revision using competitor's device was performed.At last study follow up, the patient's outcome was excellent.
 
Manufacturer Narrative
H3, h6: it was reported that on literature review: "can hip resurfacing be safely revised with short-stem total hip arthroplasty? a case series of six patients", 1 patient operated in 2004 with bhr presented limited hip mobility and pain at hip rotation since 2016.After diagnosis with acetabular loosening and acetabular bone defect, a revision using competitor's device was performed.At last study follow up, the patient's outcome was excellent.As no device part and batch numbers were provided for investigation, a complaint history review, manufacturing record review, device labelling / ifu, historic escalation actions review could not be performed.If more information is received, this investigation will be reopened.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.Without clinically relevant patient-specific supporting documentation, a thorough medical investigation could not be performed.The images provided in the article have been interpreted within the text; therefore, no further analysis of the images are required.The root cause and/or patient outcome beyond that which was documented in the article could not be confirmed nor concluded; therefore, no further medical assessment is warranted at this time.Should clinical documentation become available in the future, a thorough medical assessment may be rendered at that time.Based on the information provided we cannot confirm or further investigate the reported complaint, our investigation remains inconclusive, and a definitive root cause cannot be determined.Additionally, specific factors known to contribute to the alleged fault cannot be provided due to the insufficient information.If additional information becomes available in the future, this case will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN BIRMINGHAM HIP RESURFACING (BHR) CUP (UNKN TYPE)
Type of Device
PROSTHESIS,HIP,HEMI-,TRUNNIONBEARING,FEMORAL,METALPOLYACETAL
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14318119
MDR Text Key292062021
Report Number3005975929-2022-00189
Device Sequence Number1
Product Code JDH
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Literature
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-