(b)(4).Batch # x9402l.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Batch number was received and dhr is pending review.When the review is completed, a supplemental medwatch will be sent with a summary of the evaluation.Additional information provided : the device was used on the blood vessel in the right hemicolon.No bleeding or no damage of the tissue was observed.The clip was loose, so it was removed.Another el5ml was used to complete the case.The patient's condition is stable.Did device fire malformed clips? no further information is available.Did device fire scissored clips? no further information is available.Did clip not hold on tissue or vessel? yes.
|
(b)(4).Date sent: 6/8/2022.Investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the el5ml device was received with no damage to the external components.Upon cycling, the instrument was noted to be empty and locked out.The instrument is designed to lockout after all the clips have been fired; therefore a potential cause for the customer reported experience is the firing of all of the clips, as a result, the instrument could no longer be fire due to the activation of the lockout mechanism.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, no anomalies were found.The instrument has an orange indicator that appears on the top of the handle as a reference for the user as to the number of clips remaining.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: when the 13th clip is fired, an orange bar will begin to appear in the indicator window on top of the device handle.The orange bar fills the indicator window when the final clip is fired.The event described could not be confirmed as the device was returned empty.
|