Model Number BI70002000 |
Device Problems
Device Damaged Prior to Use (2284); Communication or Transmission Problem (2896); Physical Resistance/Sticking (4012); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/21/2022 |
Event Type
malfunction
|
Event Description
|
Medtronic received information regarding an imaging system used right before a sacroiliac and thoracolumbar procedure.It was reported that during the first use of the imaging system, the 1d mouse would not sync to the mobile viewing station (mvs), and the light would not come on.The dongle also would not seat properly causing the estop to come on.No impact on patient outcome.No delay in surgical time.
|
|
Manufacturer Narrative
|
The system was serviced in the field and it was found that standoffs were loose at dongle connection.They were refastened.Dongle and setup mouse were reseated.Performed system checkout and the system returned to service.If information is provided in the future, a supplemental report will be issued.
|
|