It was reported that there was an issue with pl574t - challenger ti-p sm-ligat.Clips 12 cartr.According to the complaint description, as the surgeon pulled the applier out of the port, it was noted that the "root of the magazine was floating".When the magazine was examined, it was found that the pin was missing from the cartridge.It was unknown whether a fragment remained in situ.The adverse event is filed under aag reference (b)(4).
|
Investigation: the product was not returned, and a thorough evaluation could not be performed.Batch history review: due to the fact that no lot number was provided, a review of the device history records for the complained device is not possible.The review of risk assessment revealed that the overall risk level ((3)5 severity x (1)5 probability of occurrence) according to din en iso 14971 is still acceptable.Explanation and rationale: without the complaint sample available, we assume that the latch on the proximal end of the cartridge broke off.Improper handling can be mentioned as a possible root cause.If the cartridge is handled with rough forces, it is possible that the latch breaks off during the removal from the packaging or during the assembly onto the shaft.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the product is returned in the future, the complaint will be re-opened and an investigation performed at that time.Based on the investigation results a capa is not required.
|