SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number CA1700/000/000JP |
Device Problem
Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that an abnormal sound was heard from the product.No additional information.
|
|
Manufacturer Narrative
|
Other, other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#(b)(4).No problem or issues were identified during the device history record review.Three pictures and a sample was returned for evaluation.During visual inspection, a hole was found.The returned sample was visually inspected under normal conditions of illumination according to site inspection procedure.The sample present a tier damage (hole) in the circuit which confirms the failure mode reported.During functional testing, leak testing was on performed on the breathing circuit and air leakage was confirmed.There was a humming sound of air leaking from the hole.The reported event was confirmed.However, the root cause cannot be associated with the manufacturing process since the failure reported could not be reproduced using the tools from the molding and assembly process.Therefor the damage (hole) was created after the product was manufactured.No corrective actions were taken since the investigation did not reveal that the defect was caused during the manufacturing process.
|
|