BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU BD INTIMA II¿ IV CATHETER PRN ADAPTER; INTRAVASCULAR CATHETER
|
Back to Search Results |
|
Catalog Number 383019 |
Device Problem
Burst Container or Vessel (1074)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/20/2022 |
Event Type
malfunction
|
Event Description
|
It was reported bd intima ii¿ iv catheter prn adapter burst, resulting in leakage.The following information was provided by the initial reporter: "there was a tube burst during the ct scan.".
|
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported bd intima ii¿ iv catheter prn adapter burst, resulting in leakage.The following information was provided by the initial reporter: "there was a tube burst during the ct scan.".
|
|
Manufacturer Narrative
|
H6: investigation summary: 45psi leakage test the retained sample of this batch , the test is qualified , no leakage is found.A device history review was conducted for lot number 1293557.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.Additionally, our engineers were able to evaluate the expanded tubing in the photographs submitted by the facility.Based on their evaluation and your description of the event, they were able to determine that the most likely root cause for this event is the forcing of fluid through the device during injection.The bd intima-ii is an infusion only device and is not rated for high pressure injections.
|
|