BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU BD INTIMA II¿ IV CATHETER PRN ADAPTER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383019
Device Problem Burst Container or Vessel (1074)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/20/2022
Event Type  malfunction  
Event Description
It was reported bd intima ii¿ iv catheter prn adapter burst, resulting in leakage.The following information was provided by the initial reporter: "there was a tube burst during the ct scan.".
 
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported bd intima ii¿ iv catheter prn adapter burst, resulting in leakage.The following information was provided by the initial reporter: "there was a tube burst during the ct scan.".
 
Manufacturer Narrative
H6: investigation summary: 45psi leakage test the retained sample of this batch , the test is qualified , no leakage is found.A device history review was conducted for lot number 1293557.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.Additionally, our engineers were able to evaluate the expanded tubing in the photographs submitted by the facility.Based on their evaluation and your description of the event, they were able to determine that the most likely root cause for this event is the forcing of fluid through the device during injection.The bd intima-ii is an infusion only device and is not rated for high pressure injections.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INTIMA II¿ IV CATHETER PRN ADAPTER
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
no.1 liangpu street
suzhou industrial park
suzhou 21512 6
CH  215126
Manufacturer (Section G)
BECTON DICKINSON MEDICAL DEVICES CO., LTD. SUZHOU
no.1 liangpu street
suzhou industrial park
suzhou 21512 6
CH   215126
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14385112
MDR Text Key292191677
Report Number3014704491-2022-00190
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number383019
Device Lot Number1293557
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial