The mayfield skull clamp (a1059) was returned for evaluation: failure analysis - the investigation of the returned device showed that the reported complaint was confirmed from the evaluation.The locking mechanism had some movement, the screw of the torque screw was damaged, the thread of the base and a few of small parts were worn off, the unit needs to be marked with instance number, and the index knob is an old version.Root cause - complaint confirmed via inspection of the unit.The screw of the torque screw was damaged, and the thread of the base was worn.Unit required replacement of worn and damaged parts.Probable root cause is rough handling of the unit and routine wear.This unit is beyond integra¿s 7 years recommended life cycle (2009).No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.
|