Model Number A9MRLVG24580TCS |
Device Problems
Use of Device Problem (1670); Insufficient Information (3190)
|
Patient Problem
Fall (1848)
|
Event Date 04/10/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is in progress.The conclusions will be provided within follow up report once the investigation is completed.
|
|
Event Description
|
Following the information provided the patient fell from the atmosair 9000 mattress, which was placed on arjo minuet bed.At the time of event the bed backrest was raised, the patient slid down the mattress and fell on the floor.No injury was sustained.The mattress was inspected and no malfunction was identified.
|
|
Manufacturer Narrative
|
The mattress has an inclined foot end, to decrease the pressure applied to the heels, which might also influence the patient's position on the bed when placed incorrectly.The instructions for use for atmosair 9000 (407391-ah) includes the following information related to patient falls: - "patient migration - specialty surfaces have different shear and support characteristics than conventional surfaces and may increase the risk of patient movement, sinking and/or migration into hazardous positions of entrapment and/or inadvertent bed exit." - "head of bed elevation - keep head of bed as low as possible to help prevent patient migration." - "position mattress on bed frame with logo facing up and product information tags at foot end of bed." it also includes the symbol marking foot end of the bed.Based on the collected information and the assessment made by the arjo representative, the root cause of the complained scenario is considered use error.The device was evaluated after the event and no malfunction was found.It is unknown what was the backrest elevation angle used or whether there was anything on the mattress that could reduce friction between the patient's body and mattress surface (e.G.Patient's clothing material or sheets).Arjo device was used for a patient treatment when the event occurred and from that perspective it played a role in the event.No malfunction was found during the device evaluation.This complaint is deemed reportable due to patient's fall from the device.
|
|
Event Description
|
Following the information provided the patient fell from the atmosair 9000 mattress, which was placed on arjo minuet bed.At the time of event the bed backrest was raised, also the egress assist side rail of the bed was raised.The patient slid down the mattress and fell down to the floor.No injury was sustained.The patient was picked up off the floor with a maxi move floor lift and put back to bed.At that time it was noted that the mattress was placed reversely on the bed - the foot end of mattress was at the head end of the bed.The mattress was inspected and no malfunction was identified.The arjo representative recommended the customer to use higher side rails.
|
|