It was reported to philips that the device experienced ecg cannot detect discharge issue.There was no report of patient involvement.The customer requested that an authorized field service engineer (fse) be dispatched to the customer site to evaluate the noted issue.Upon evaluation of the device, the reported issue was confirmed and the cause was traced to a faulty capacitor and processor pca.Based on the conclusion of the evaluation, it was determined that this was a malfunction of the capacitor and processor pca.Multiple parts were replaced to resolve the noted issue.The device passed all performance assurance tests and was placed back into use with the customer.The device remains at the customer site.There is no indication of a systemic problem.No further investigation or action is warranted.We are unable to determine the root cause of the event as multiple parts were required for replacement.
|