• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME, LINE-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME; DERMATOME, LINE-POWERED Back to Search Results
Model Number N/A
Device Problems Loose or Intermittent Connection (1371); Failure to Power Up (1476); Failure to Cut (2587); Noise, Audible (3273)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 05/09/2022
Event Type  malfunction  
Event Description
It was reported that the power application to the unit is not effective with activating the unit; the power did not switch on immediately accompanied by a clinging sound (as if something is loose on the inside).The dermatome setting for the graft was on the thinnest option, but still made a cut into the sub-cutaneous fat.The event timing was during surgery.The patient needed stitches to close the wound created by the device.
 
Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).A follow up/ final report will be submitted once investigation is complete.
 
Event Description
There is no additional event information available.
 
Manufacturer Narrative
Following sections were updated or corrected : b4, b5, d4, d9, g6, g3, h1, h2, h3, h4, h6, h10.Review of the most recent repair record identified no repairs related to the reported event.The reported harvesting issue could not be replicated.The device passed final testing and extensive quality protocol; no device problem found.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME, LINE-POWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key14427489
MDR Text Key291937212
Report Number0001526350-2022-00464
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024376994
UDI-Public(01)00889024376994(11)181003(10)6418744
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-