Model Number MHD8S
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2022
Event Type  malfunction  
Event Description
It was reported that a heparinized saline leakage was observed at the time of priming before use.The exact location of the leakage is unknown.There were no patient complications reported.
 
Manufacturer Narrative
The device evaluation is anticipated.However, the complaint cannot not be confirmed without the completion of a product evaluation.A supplemental report will be forthcoming with the evaluation and device history results when received.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Manufacturer Narrative
Heparinized saline is used per hospital protocol to maintain patency in the pressure line/catheter and in this case was not going to be used as a treatment for the patient.There was no loss of therapeutic medication, which in accordance with our decision rationale matrix does not meet the criteria for an mdr reportable event.A product evaluation was completed.The reported event of leakage was confirmed.As received, flotrac housing male luer had been completely broken.The broken luer was found to be jagged.Leakage was most likely due to broken luer.No leakage was found from rest of the unit during leak test.No other damage was observed from the kit.Per evaluation findings, device was confirmed for broken luer.This is not a reportable event.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Manufacturer Narrative
It was found within our procedure and decision rationale documentation that this event was reported in error.This is not a reportable event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOTRAC SENSOR
Type of Device
TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
Manufacturer Contact
jonathan diaz
1 edwards way
irvine, CA 92614
MDR Report Key14438483
MDR Text Key292071351
Report Number2015691-2022-05759
Device Sequence Number1
Product Code DRS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043065
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 06/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date08/04/2023
Device Model NumberMHD8S
Device Lot Number63870514
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-