The customer reported that end user immediately noticed unusual co2 trace pattern/waveform on the anesthesia machine during hip replacement operation.The patient was transferred to a different machine in good time, and no harm is associated with the incident.However, it did cause stress to the team.Upon checking, the medisorb¿ multi-absorber original, disposable cannisters seemed to be too full or unable to shake or did not seem to have the movement as they normally do.This was proven by using a different product (unspecified).Also, general electric was called to test their equipment, and this passed and was in good working order.
|