MEDTRONIC NAVIGATION, INC FRAME, 9732353, PERC REF, CROSS PIN; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,
|
Back to Search Results |
|
Model Number 9732353 |
Device Problems
Imprecision (1307); Mechanics Altered (2984)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/25/2022 |
Event Type
malfunction
|
Event Description
|
Medtronic received information regarding a navigation device being used during a spinal procedure.It was reported that there was a 5-10mm inaccuracy during navigation.Auto-registration was used.The ball tip plana probe was being used when the inaccuracy was detected.Additionally, a universal drill guide (udg) and awl tip taps were being used.The percutaneous pin frame was being used.The frame was located at the right posterior superior iliac spine (psis).The percutaneous pin frame was loose and would not tighten properly.After images were taken, the images did not match any instruments.After the inaccuracy was discovered, a midline incision was made and a spinous process clamp was placed on the l4 spinous process.Another image was taken and the procedure was completed.The procedure was delayed by less than one hour.There was no impact on patient outcome.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|