MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP; RSP, PRIMARY HUMERAL MONOBLOCK, 7MM X 108

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP; RSP, PRIMARY HUMERAL MONOBLOCK, 7MM X 108 Back to Search Results
Model Number 510-00-007
Device Problems Loose or Intermittent Connection (1371); Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Date 05/04/2022
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event Description
Revision surgery - due to loosening.
 
Manufacturer Narrative
The reason for this revision surgery was reported as loosening.The previous surgery and the surgery detailed in this event occurred 7 years and 3 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements and showed no non-conforming material reports (ncmr) associated with the main contributor component listed in the complaint.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to loosening.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, prolonged overhead activities, inadequate soft tissue support, patient activities.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP
Type of Device
RSP, PRIMARY HUMERAL MONOBLOCK, 7MM X 108
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key14458807
MDR Text Key292202607
Report Number1644408-2022-00682
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888912024808
UDI-Public(01)00888912024808
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100741
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2020
Device Model Number510-00-007
Device Catalogue Number510-00-007
Device Lot Number978F1051
Was Device Available for Evaluation? No
Date Manufacturer Received05/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/26/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
508-00-036 LOT 856C1095; 508-36-103 LOT 871C1082
Patient Outcome(s) Required Intervention;
Patient Age69 YR
Patient SexFemale
-
-