Model Number CLV-S190 |
Device Problems
Electrical /Electronic Property Problem (1198); Output below Specifications (3004)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/21/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The referenced device was returned to the repair center for evaluation.The customer¿s reported issue was confirmed; the e200 error was attributed to a defective/faulty filter turret and the front panel indicator flashed.The root cause of the faulty turret cannot be confirmed.A device has not been repaired in the past year.The legal manufacturer performed a review of the device history records for the concerned device and no abnormalities were found.In general, the customer is required to check the function of all devices used prior to a procedure and as a preventative measure, a suitable replacement device must be provided during an application.Olympus will continue to monitor complaints for this device.
|
|
Event Description
|
Olympus (osh) was informed by the user facility via service inquiry that the customer visera elite xenon light source displays a reported ¿e200, front panel indicators flash and the lamp is dark¿.The event was found during reprocessing.There was no delay as the scheduled laryngoscopy was completed using another similar light source.No death, injury or harm was reported to olympus.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to correct information provided on the initial report.The following sections were corrected: g2, specify other as ¿china¿.It has been over 8 years since the subject device was manufactured.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|