NIDEK CO. LTD. YC-200 S PLUS; YAG LASER Back to Search Results
Model Number YC-200
Device Problem Unexpected Therapeutic Results (1631)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 05/06/2022
Event Type  malfunction  
Event Description
Nidek inc received a telephone call from the customer to report that they observed pitting lens when fired the laser for the first time.Focus shift setting was 250 posterior.The pitting lens happen last week, and the doctor informed the technician at the customer site to report pitting lens to nidek inc on (b)(6) 2022, and nidek inc received a telephone call on 05/09/2022.No injury was reported from the customer.Device evaluation is pending.Nidek inc.Considers pitting lens issue a reportable event as it is an undesirable condition and has the potential to cause or contribute to a serious injury if the issue were to recur.Technical background: the ophthalmic yag laser system yc-200 consists of a slit lamp and the yag laser and is indicated for the performance of posterior capsulotomy, pupillary membranectomy, iridotomy (hole in the iris) and selective laser trabeculoplasty.This device uses passive q-switching for laser oscillation.With passive q-switching, the device first accumulates sufficient energy within the laser cavity then emits laser pulses of high peak power for extremely short durations by use of the q-switched optical element which changes between transparent and opaque according to the accumulation of energy.Yc-200 device is equipped with yag mode function only.Yc-200 s plus device is equipped with yag and slt mode functions.
 
Manufacturer Narrative
As of 05/19/2022, the device evaluation is pending.Nidek inc senior product marketing manager and senior sales & marketing manager are communicating with the user facility to schedule a visit.Follow-up mdr will be submitted once nidek inc receives further information from the user facility.
 
Event Description
Please see initial mdr submitted on 05/19/2022.
 
Manufacturer Narrative
Nidek inc.Determined that the cause of customer reported issue "pitting lens" was inconclusive.Nidek inc field service engineer (fse) powered on system and found in working order.Shot counter = 1,298.Measured 2.8mj at 3.0mj setting.Measured 4.7mj at 5.0mj setting.Optical breakdown = 1.5mj.Beam coincidence checked ok.Fse performed following service activities as preventive measures: o performed complete inspection of the laser system.O auto-calibration of the pfn voltage = 465vdc.O cleaned objective lens and oculars.O for good measure, calibrated laser energies: 0.3mj, 0.5mj, 1.0mj, 3.0mj, 5.0mj, 7.0mj, and 10.0mj in single, double, and triple burst modes.O verified: yag/aiming coincidence and offset focus shift is proper for 0 - 500 microns anterior and posterior.Yag energy output is in spec 0.2 - 10.0mj single burst, 0.4 - 20.0mj double burst, 0.6 - 30.0mj triple burst.Optical breakdown = 1.4mj.System operational at all settings and modes.All test shots and beam profile checked ok.The system was tested and inspected.Fse verified that the system is operational.The reported unit passed final inspection per inspection distribution acceptance record #idar-2021-0269, and installation and in-service training was performed properly per ins-2021-0144.There was no supporting evidence to point to machine, man, method/process and material as the cause of the reported complaint issue.As a precaution,rg-093-2 yc-200/ yc-200 s plus quick reference guide will be sent out to the customer to remind customer of safety information in yc-200 operator's manual to mitigate against potential occurrence of pitting lenses.Per device evaluation, fse verified that the system is operational.Based on the historical data review of 1 year, there are 6 related similar complaints but no device malfunction found, therefore no trend identified.Investigation is closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
YC-200 S PLUS
Type of Device
YAG LASER
Manufacturer (Section D)
NIDEK CO. LTD.
34-14 hiroishi
gamagori, aichi 443-0 038
JA  443-0038
Manufacturer (Section G)
NIDEK CO. LTD.
34-14 hiroishi
gamagori, aichi 443-0 038
JA   443-0038
Manufacturer Contact
todd milholland
2040 corporate court
san jose, CA 95131
4084686460
MDR Report Key14459549
MDR Text Key300379464
Report Number0002936921-2022-00005
Device Sequence Number1
Product Code HQF
UDI-Device Identifier04987669101452
UDI-Public(01)04987669101452
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203130
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberYC-200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/28/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-