MEDTRONIC EUROPE SARL AZURE XT DR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE
|
Back to Search Results |
|
Model Number W2DR01 |
Device Problems
Signal Artifact/Noise (1036); Failure to Capture (1081); Device Contamination with Body Fluid (2317)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/28/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that four months post implant, the right atrial (ra) lead was found to have dislodged.The lead was repositioned and remains in use.During the replacement procedure, the lead was reconnected to the implantable pulse generator (ipg) after positioning, and a loss of capture and noisy electrograms (egm) was observed.It was suspected that blood had gotten into the ports, and despite being flushed with saline multiple times, the same issues were observed upon testing.The device was subsequently explanted and replaced.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|