Model Number 27003
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2022
Event Type  malfunction  
Manufacturer Narrative
The astral device was returned to resmed for an investigation.Review of the device data logs confirmed the reported complaint and revealed an error message (sf180) related to a battery charger fault.The internal battery was replaced to address the issue.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the reported complaint was due to an isolated component failure within the device battery assembly.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed an internal battery degraded warning alarm.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - AMER
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
Manufacturer (Section G)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
Manufacturer Contact
michelle hughes
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
MDR Report Key14465583
MDR Text Key298050881
Report Number3004604967-2022-00359
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00619498270033
UDI-Public(01)00619498270033(11)200804(10)1472577
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172875
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27003
Device Catalogue Number27003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2022
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date04/25/2022
Date Manufacturer Received04/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-