It was reported that, after left bhr surgery was performed on (b)(6) 2009 due to osteoarthritis, the patient experienced osteolysis, buttock pain and elevated metal levels.This adverse event was treated with a revision surgery on 11-may-2021, in which joint fluid, synovial hypertrophy and osteolysis of the femoral neck were found.The patient's current condition is unknown.
|
Investigation conclusions).Results of investigation: it was reported that left hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the acetabular cup and the femoral head was performed using batch numbers to evaluate patterns of repeated failures or defects over the lifetime of the product and using part numbers and the reported failure modes to evaluate patterns of repeated failures or defects in a timeframe prior 12 months as of the complaint aware date.No other complaints were identified to involve this batches.Other similar complaint was identified for the part numbers and the reported/related failure mode.However, as the devices are no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions required.The available medical documents were reviewed.With the information provided the clinical root cause of the reported elevated metal levels cannot be confirmed.The reported buttock pain may be consistent with the synovial hypertrophy and osteolysis; however, it cannot be concluded the reported clinical reactions were associated with a mal performance of the implant or implant failure.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Based on the available information we can confirm the reported complaint, however without further information our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.Corrected data: h3, h6 (health effect - clinical code).
|