Model Number 27074
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2022
Event Type  malfunction  
Event Description
During resmed evaluation, an astral device displayed an internal battery degraded alarm and an error message (sf74) related to the software.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
The internal battery was replaced and device reset to address the issues.The device was serviced and fully tested before it was returned to the customer.Based on all available evidence and complaint investigations of a similar nature, an investigation determined that the internal battery degraded warning alarm was due to an isolated component failure within the device battery assembly while sf74 was due to the software.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 DBL KIT FRA
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
Manufacturer (Section G)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
Manufacturer Contact
michelle hughes
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
MDR Report Key14477059
MDR Text Key298051170
Report Number3004604967-2022-00362
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K172875
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27074
Device Catalogue Number27074
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2022
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date04/25/2022
Date Manufacturer Received04/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/20/2015
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-