Catalog Number 03666794001
Device Problems Circuit Failure (1089); Failure to Power Up (1476); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/29/2022
Event Type  malfunction  
Manufacturer Narrative
The meter was requested to be returned for investigation.The investigation is ongoing.
 
Event Description
There was an allegation of a melted battery cover for a coaguchek xs meter.The patient's meter would not power on.The patient changed the batteries and reset the meter.The back of the battery door appeared to be melted and one of the batteries showed signs of corrosion.
 
Manufacturer Narrative
The investigation reviewed the picture of the meter's battery cover and confirmed there was residue from a leaked battery.The meter was provided for an investigation.The investigation could not turn on the meter.The device data and fault memory were read out.A visual examination was performed.The batteries used by the user were not returned.The deposits from a leaking battery were found in the battery compartment.This also affects the battery contact, which was temporarily interrupted as a result.Per product labeling, "if you store the meter for a period of time, remove the batteries." the investigation determined the issue was consistent with improper meter handling.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key14488686
MDR Text Key294037972
Report Number1823260-2022-01469
Device Sequence Number1
Product Code GJS
UDI-Device Identifier04015630945689
UDI-Public04015630945689
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received07/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-