OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-90
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
One device was returned for investigation.The returned sample was visually inspected, and the device was noted to have a cracked tank cover and an outdated printed circuit board (pcb) and power switch were noted.The investigator started with a visual inspection then filled tank with water, attached temp check, plugged in line cord, and turned on the power switch.Then ran on the safety/electrical test which it failed.To use the test to help diagnose the problem the main components are unplugged from the pcb one by one, and the test is run again each time a component is removed.The customer's complaint was confirmed, the device failed the safety/electrical test.In this case it was the heater that caused the failure.They were unable to determine the cause of the failure.No action taken due to the age and condition of the device.It was deemed beyond economical repair and will be scrapped.A manufacturing dhr review was not performed because the results of the complaint investigation do not indicate a problem with the manufacture of the device.No causes or potential causes of the customer's reported problem were found during the review of service and repair records.
 
Event Description
It was reported that the fluid warming device failed an electrical safety test.No patient injury was reported.No other information has been provided at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14496625
MDR Text Key292806462
Report Number3012307300-2022-09615
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2022
Was the Report Sent to FDA? No
Date Manufacturer Received04/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-